Page 128 - The Veterinary Laboratory and Field Manual 3rd Edition
P. 128

The selection, use, maintenance and quality control of laboratory equipment and supplies   97



































                Figure 2.66  Pipette graduated. Photo: Willy   Figure 2.67  Pipette volumetric close up. Photo:
                Schauwers, Provincial Institute for Hygiene,   Willy Schauwers, Provincial Institute for Hygiene,
                Antwerp, Belgium.                        Antwerp, Belgium.


                HandLInG oF PIPEttES                         contact with the inclining inner surface of
                Pipettes calibrated ‘to deliver’ (‘TD, Ex’).  the receiving vessel
                                                           •  the pipette is held nearly vertical, the tip is
                •  Filling:
                                                             placed in contact with the inclining inner
                  •  a pipette bulb or pipettor (manual pi-pump   surface of the receiving vessel
                     or electric/battery operated pipette-aid) is   •  when the meniscus comes to a rest in the
                     attached to the upper end and the liquid   tip, the waiting time of 5 s begins (class
                     is suctioned up into the pipette ±5 mm   AS only)
                     above the marking on the stem above the   •  after the waiting time, draw the pipette
                     centre bulb                             upwards along the inner wall of the receiv-
                  •  wipe off any remaining liquid with a tissue   ing vessel through a distance of about
                     at the outside of the tip               10 mm (Figure 2.67)
                  •  set the meniscus                      •  pipettes should be rinsed with tap water
                  •  wipe off any drop of fluid adhering to the   (by using a water jet pump) immediately
                     tip.                                    after use.

                •  Delivering:
                  •  hold the pipette vertically and deliver
                     the liquid with the tip of the pipette in







       Vet Lab.indb   97                                                                   26/03/2019   10:25
   123   124   125   126   127   128   129   130   131   132   133