Page 129 - The Veterinary Laboratory and Field Manual 3rd Edition
P. 129

98  Willy Schauwers

            Pipettes calibrated ‘to contain’ (‘TC, In’).  instructions of the manufacturer. For how to
                                                     use mechanical pipettes see section ‘Automatic
            •  Filling:                              mechanical pipettes and dispensers’.

              •  aspirate the liquid to the mark     WorkInG WItH PIPEttE aIdS
              •  hold the pipette horizontally and carefully   Pipetting by mouth is not allowed. The use of
                 wipe off with a tissue.             pipetting aids is mandatory. Pipetting aids can
                                                     be manual or motorized.
            •  Delivering:                             Manual pipetting aids (pipette fillers) use a
                                                     valve system to create pressure (Figure 2.68).
              •  to empty capillaries blow out the liquid
                 with a pipetting aid, and rinse two to three   •  The pipetting aid (pipette filler) is attached to
                 times with the diluting medium (required   the top of the pipette.
                 due to calibration ‘to contain’).   •  Press ‘A’ to squeeze the ball and to create
                                                       negative pressure.
            Mechanical, air-displacement  or positive dis-  •  Press ‘S’ to fill the pipette just above the
            placement pipettes with disposable tips are   mark.
            frequently used to measure small volumes.   •  Press ‘E’ to drain the pipette.
            They are available in a variety of volumes, rang-
            ing from 1 to 5000 µl. Used tips are disposed of   Motorized pipettors (Figure 2.69) are useful
            directly into disinfectant using the ejector mech-  for longer pipetting series. The pipette aid is
            anism of the pipette. The micropipettes usually
            have two positions of the plunger operated by
            thumb. The first position is used to pick up the
            sample and the second to expel the sample from
            the tip into a tube or well. Micropipettes must
            be calibrated and maintained according to the























            Figure 2.68  Pipette filler. Photo: Willy Schauwers,   Figure 2.69  Pipette filler electric. Photo:
            Provincial Institute for Hygiene, Antwerp, Belgium.   Willy Schauwers, Provincial Institute for Hygiene,
                                                     Antwerp, Belgium.







       Vet Lab.indb   98                                                                   26/03/2019   10:25
   124   125   126   127   128   129   130   131   132   133   134