Page 224 - Adams and Stashak's Lameness in Horses, 7th Edition
P. 224

190   Chapter 3


              Kilovoltage determines the energy of X‐rays being   X‐ray  detector being used. Collimating the primary beam
            produced. Kilovoltage potential should range from 70   is a safety practice that must be monitored for each
  VetBooks.ir  when a portable X‐ray machine is used. A 10% increase   Limiting the size of the primary X‐ray beam is a major
                                                               exposure and must be reset if the cassette size changes.
            to 90 for equine limb radiographs on mature horses
                                                               factor in reducing scatter radiation, which keeps radia­
            or decrease in kilovoltage potential effectively doubles
            or halves the radiation exposure.                  tion exposure to personnel as low as possible.
              The focal spot‐film distance (FFD) is the distance   Collimation also improves image quality, particularly
            from the X‐ray tube focal spot to the detector. The focal   in large body parts such as the caudal cervical spine.
            spot location is usually marked on the outside tube   Fixed cylinders or cones and adjustable light‐beam col­
            housing; if it is not, the approximate center of the tube   limators are available (Figure 3.2). The disadvantage
            housing can be used for its location. A constant FFD is   of fixed primary beam restrictors is that they do not
            necessary to minimize improper exposures because the   conform to different cassette sizes.  Adjustable light‐
            intensity of X‐rays that expose the film is inversely   beam collimation is recommended for equine radiogra­
              proportional to the square of the FFD.  Therefore, a   phy  and  is  standard  to many  X‐ray  machines. They
            small difference in the FFD can dramatically change the   come with an internal light source, preferably 40 W or
                                                                      48
            exposure on the X‐ray film.                        greater,  so the limits of the primary X‐ray beam are
              Some method of measuring the FFD before each expo­  projected on the X‐ray cassette/detector as visible light.
            sure should be employed. Measuring devices with       The type of X‐ray machines best suited for equine
              converging light beams at the correctly set FFD are easy to   practice depends on the type of practice: out of hospital
            use and less cumbersome than a measuring tape or string.   versus in hospital. The features of the X‐ray machines
            The FFD for equine limb radiography should be between   must be matched to fit different practice situations.
            36 and 40 inches (85 and 100 cm) and no less than   Some compromises or trade‐offs must be made, e.g. less
            24 inches (60 cm). When the FFD is less than 24 inches, the   milliamperage or kilovoltage potential for more porta­
            object is magnified, and spatial resolution is reduced. FFDs   bility. An ideal X‐ray machine for equine radiography
            greater than 40 inches can be used if exposure times are   has the following features:
            not excessively long. When making radiographs of larger   1.  Easily and quietly movable, with a tube head that can
            body parts such as necks, it is important to remember that
                                                                  extend to the floor or ground surface
            the distance needs to be set from the X‐ray tube to the   2.  Adjustable milliamperage and kilovoltage potential
            detector, not to the skin surface, a concept that is more
                                                                  setting
            easily forgotten  when  the  plate is  not  well  seen  by the   3.  Electronic,  two‐step  timer  capable  of  accurate
              person holding the X‐ray generator.
              Collimation is the process of limiting or restricting     exposure times of 0.1 second or faster
            the  primary X‐ray  beam to  the appropriate  size to   4.  Line voltage compensator and compensation meter
            cover the anatomical region of interest within the



































                                                               Figure 3.2.  Adjustable light‐beam collimator. The primary beam
                                                               is limited by adjustable leaves in the collimator. A light source within
            Figure 3.1.  Control panel for a 30‐mA portable X‐ray machine.  the collimator defines the limits of the primary beam (white arrows).
   219   220   221   222   223   224   225   226   227   228   229