Page 55 - Quality control of pharmaceuticals (07-PA 704)
P. 55

Example (4):
Comment on the accuracy and precision of the following data provided by two
analysts in the same laboratory using exactly the same method.

Exp No.  Analyst 1  Analyst 2
    1       99.7       99.8
    2       98.9      101.7
    3       99.1      102.8
    4      100.1      100.2
    5       99.9       98.9
    6       99.1      101.0

Solution
   1. Calculation of t calcu = 2.16

t-/theor = 2.228 (From t-table)
   2. Calculation of variances S2

Fcalc = 1.958 /0.559 = 3.502
Ftheor = 5.05 (From F table)
Acceptance Limit: calculated t and F should be lower than the theoretical
Conclusion:
Since the t calc less than t theor: both analysts' mean results have the same
accuracy. Since the Fcalc less than F theor: both analysts have the same precision.
   50   51   52   53   54   55   56   57   58   59   60