USABILITY
                                                      EFFICACY








                       USABILITY TESTING REPORT FOR MEDICAL DEVICE.

                   TIM TRANSPARENCE ET INFORMATION DU MÉDICAMENT.





                                            (According to ISO IEC 62366:2007) Jun 2013




                 ABSTRACT


                 PURPOSE: Comparative study to evaluate the usability of 3 containers (COMOD®, Control grip™, Easy
                 grip®).


                 METHODS: 40 participants from 18 to 78 whose 24 participants aged of more than 60, face to face
                 interviews with a questionnaire after test of pairs of bottles (water / empty). 3 containers are tested
                 (COMOD®, Control gripTM, Easy grip®).


                 STUDY CRITERIA: Handling, pumping mechanism, administration into the eyes.

                 RESULTS: The participants considered the handling of Easy grip® easier or as easy compared with the
                 two other devices (100% vs COMOD®, 90% vs Control gripTM).
                 The activation of the pumping mechanism of Easy grip® is easier or as easy compared with the two
                 other devices (95% vs COMOD®, 90% vs Control gripTM).
                 The administration into the eyes is easier or as easy with Easy grip® compared with the two other
                 devices (90% vs COMOD®, 90% vs Control gripTM).
                 Easy Grip® allows a natural positioning of the fingers thanks to the two tabs


                 CONCLUSIONS: Easy Grip® allows a natural positioning of the fingers thanks to the two tabs com-
                 pared to handling of Easy Grip® is easier pr as easy compared to COMOD® and Control Grip™.