Page 102 - Catalogue - cuvettes and solid-state calibration standards
P. 102
NIST 2034 equivalent – Holmium Perchlorate UV/VIS wavelength accuracy
Firefly Scientific's equivalent of the NIST SRM 2034 calibration reference standard fulfills the requirements of the European
Pharmacopoeia (EP). Please note that is must be recalibrated annually. If you prefer a product that doesn't require recalibration,
we recommend the FF-WAV-1-UV/VIS.
Part Number Description Price
FF-NIST-2034 NIST 2034 compliant Holmium perchlorate wavelength accuracy filter (240-650nm) 510.53
Capability:
Calibration of the wavelength accuracy in the 240-650nm range
Specifications:
Contents: Holmium Perchlorate (4% w/v) in 10% Perchloric Acid (v/v). This corresponds to the EP requirement
for 40g/L Holmium Oxide in Perchloric Acid containing 141g/L HCl04
Wavelength peaks: 240.8, 277.8, 287.1, 361.1, 416.4, 451.2, 485.2, 536.5, 650.5nm
External dimensions: 12.5 x 12.5 x 45.0mm
94