Page 25 - 27766-LASTPAGES
P. 25

Guidance and Manufacturer's Declaration – Electromagnetic Immunity  14  Guidance and Manufacturer's Declaration – Electromagnetic Immunity
 The Product is intended for use in the electromagnetic environment specified below.   The Product is intended for use in the electromagnetic environment specified below.
 The customer or the user of the Product should assure that it is used in such an   The customer or the user of the Product should assure that it is used in such an
 environment.  environment.

 IEC 60601-1-2   Compliance   Electromagnetic   IEC 60601-1-2   Compliance   Electromagnetic
 Immunity test  Immunity test
 Test level  level  environment - guidance  Test level  level  environment - guidance
 Floors should be wood,                                Mains power quality should be
 Electrostatic   ± 8 kV contact  ± 8 kV contact  concrete or ceramic tile. If   Voltage dips  that of a typical public low-
 discharge   floors are covered with   0% UT;   0% UT;   voltage power supply network
 (ESD)   ± 15 kV air  ± 15 kV air  synthetic material, the relative   0.5 cycle  0.5 cycle  that supplies buildings used
 IEC 61000-4-2  humidity should be at least            for domestic purposes,
 30%.                    0% UT; 1 cycle  0% UT; 1 cycle  commercial or hospital, clinical
 ± 2 kV for   ± 2 kV for   Mains power quality should be   and  and  environment. If the user of the
 power supply   power supply   that of a typical public low-  70% UT;  70% UT;  Product requires battery
 Electrical fast         25/30 cycles   25/30 cycles   charging during power mains
 transients /   lines  lines  voltage power supply network   Voltage   interruptions, it is
 bursts,   that supplies buildings used   interruptions  0% UT;  0% UT;  recommended that the
 IEC 61000-4-4  ± 1 kV for   Not Applicable  for domestic purposes,   IEC 61000-4-  250/300 cycles  250/300 cycles  Product charger be powered
 input/ output   commercial or hospital, clinic   11
 lines  environment.                                   from a separate power supply
                                                       (UPS, etc.).
 Mains power quality should be
 ± 1 kV Line-to-  ± 1 kV Line-to-  that of a typical public low-  Power frequency magnetic
 Surges  line  line  voltage power supply network      fields should be at levels
 that supplies buildings used   Rated power            characteristic of a typical
 IEC 61000-4-5  ± 2 kV Line-to-  ± 2 kV Line-to-  for domestic purposes,   frequency   30 A/m  30 A/m  public low-voltage power
 ground  ground  commercial or hospital, clinic   magnetic fields  50 or 60 Hz  50 or 60 Hz  supply network that supplies
 environment.                                          buildings used for domestic
                                                       purposes, commercial or
                                                       hospital, clinic environment.
                         Note
                         UT is the a.c. mains voltage prior to application of the test
                         level.

















 216/222  BMLPXEUDFUPRT / Rev.13 / 10-2018  A1030 000 001 00  13  13  BMLPXEUDFUPRT / Rev.13 / 10-2018  217/222                                                                       #                                                               27766-EYAL - 277766-EYAL | 7 - A | 19-01-08 | 13:40:47 | SR:-- | Yellow   27766-EYAL -
   20   21   22   23   24   25   26   27   28   29   30