Page 26 - 27766-LASTPAGES
P. 26
Guidance and Manufacturer's Declaration – Electromagnetic Immunity The recommended radiation levels of RF wireless communication
equipment specified in this paragraph must be complied with.
The Product is intended for use in the electromagnetic environment specified below.
The customer or the user of the Product should assure that it is used in such an
environment. Guidance and manufacturer’s declaration - electromagnetic emissions
IEC Compliance Electromagnetic environment - The Product is intended for use in professional healthcare facility or home healthcare
Immunity test 60601-1-2 level guidance electromagnetic environment specified below. The customer or the user of the Product
Test level should assure that it is used in such an environment.
Portable and mobile RF Emissions test Compliance Electromagnetic
communications equipment should environment - guidance
be used no closer to any part of The Product uses RF
the Product, including cables, than energy only for its internal
the recommended separation function. Therefore, its RF
3 Vrms 3 Vrms distance calculated from the RF emissions Group 1 emissions are very low and
150 kHz to 150 kHz to equation applicable to the CISPR 11 are not likely to cause any
Conducted 80 MHz 80 MHz frequency of the transmitter.
disturbances interference in nearby
induced by RF 6 Vrms in 6 Vrms in Recommended separation electronic equipment.
fields ISM bands ISM bands distance: RF emissions Class B
IEC 61000-4-6 CISPR 11
150 kHz to 150 kHz to d = 1.17 P The Product is suitable for
80 MHz 80 MHz from 150 kHz to 80 MHz Harmonic emissions Not applicable use in professional
IEC 61000-3-2 healthcare facilities or
80% AM at 80% AM at d = 1.17 P Voltage fluctuations/flicker domestic establishments.
1 kHz 1 kHz from 80 MHz to 800 MHz emissions IEC 61000-3-3 Not applicable
d = 2.3 P
Radiated RF 10 V/m 10 V/m
IEC 61000-4-3 from 800 MHz to 2.7 GHz
80 MHz to
2.7 GHz Where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is
the recommended separation
distance in meters (m).
218/222 BMLPXEUDFUPRT / Rev.13 / 10-2018 BMLPXEUDFUPRT / Rev.13 / 10-2018 215/222
#27766-EYAL - 277766-EYAL | 7 - B | 19-01-08 | 13:40:47 | SR:-- | Black 27766-EYAL - 277766-EYAL | 7 - B | 19-01-08 | 13:40:47 | SR:-- | Yellow 27766-EYAL - 277766-EYAL | 7 - B | 19-01-08 | 13:40:47 | SR:-- | Magenta 27766-EYAL - 277766-EYAL | 7 - B | 19-01-08 | 13:40:47 | SR:-- | Cyan #