 7.3 Description of User Interface   • To exit the DEMO mode, tap the RAYPEX   6
                                         ®
 WARNINGS  ®  “home” icon (Fig. 37).             NOTE
 en  For removing the charger, follow the   RAYPEX  6 has a tilt adjustable front panel with a   Servicing and repair work should only be   en
 large graphic TFT touch screen.
 sequence below:                                 carried out by service engineers trained
 •  Before removing the external charger   The main screen shows the following icons and   at the factory.
 from the power outlet, disconnect the   symbols:
 charger from the device.                     The measuring cable and the surface of the
 •  The device cannot be used while           equipment should be cleaned with a paper towel
 charging.                                    or soft cloth or wipe that has been wetted with
 Settings icon                                aldehyde-free disinfection and cleaning solution
                                              (bactericidal and fungicidal), e.g
                                              • DENTIRO  wipes, Oro Clean Chemie  AG,
                                                     ®
                                              Switzerland
 Battery Replacement    Sound icon            • TopActiv wipes, ad-Arztbedarf GmbH, Germany
 The battery compartment is located at the rear of   • SprayActiv, ad-Arztbedarf GmbH, Germany
 ®
 RAYPEX  6 and its cover is secured by screws.   Measuring cable connector and
 • Release the screws and remove the battery   lip clip/ file clip symbols  • DY wipes, DENTSPLY, France
 compartment cover.
 • Remove the battery from battery compartment
 and disconnect the battery cable plug from the   Full canal image  WARNINGS
 battery jack of RAYPEX 6.   Fig. 37  Exit DEMO mode  •  To disinfect the equipment and the
 ®
 • Insert the cable plug of the new battery into the   measuring cable, wipe the surface with
 battery jack.   Apical zoom image                a clean cloth or wipe that has been
 • Insert the battery into the battery compartment.  NOTE  wetted only slightly with a non-aggressive
 • Close the battery compartment and secure the   If the measuring cable is connected   disinfectant.
 cover with the screws.  during the DEMO mode running,   •  Do not use any liquids or sprays directly
              RAYPEX  6 will automatically exit the   on the equipment, particularly the display.
                   ®
              DEMO mode.                          •  Do not use strong alcohol to disinfect
 WARNINGS                                         surfaces.
 •  Before replacing the battery, disconnect
 the charger from the device!
 •  The used battery must be disposed of in   8  Maintenance, cleaning and   Before preparing them for reuse, remove the lip
 accordance with the local regulations        clip, file clip or touch probe from the measuring
          sterilization                       cable.
                                              The lip clip, file clip and touch probe must always
          8.1 General information             be cleaned, disinfected and sterilized before
                                              they are used. This also applies before they are
                                              used for the first time. The only way to ensure
          The RAYPEX  6 is maintenance-free and does not   that these accessories are effectively sterilized
                  ®
          have any components that need to be maintained   is to thoroughly clean and disinfect them. When
          by the user.                        doing this, follow the special instructions set
                                              out in section 8.2, "Cleaning, disinfection and
                                              sterilization".







                                 Endo Easy Efficient ®                                                                 #                                                              #  26306-EYAL - 26306-EYAL | 2 - B | 18-08-14 | 12:05:15 | SR:-- | Yellow   26306-EYAL - 26306-EYAL | 2 - B | 18-08-14 | 12:05:15 | SR:-- | Cyan   263