52                                         65                                                                                                                                        #                                                               26306-EYAL - 26306-EYAL | 2 - B | 18-08-14 | 12:05:15 | SR:-- | Yellow   26306-EYAL -


 • Then, remove the accessories from the  8.2.5  Sterilization
 disinfection bath and rinse them off thoroughly
 at least five times for one minute each time with   Use only the sterilization methods listed below;
 en  sterile, deionized water; when you rinse it, squeeze   other sterilization methods are not permitted.
 the file clip together and then release it five times.  • Steam sterilization
 • Dry the accessories by blowing them down fully   • Fractionated vacuum/pre-vacuum method (at
 with oil-free, filtered compressed air and then   least three vacuum cycles) or gravity method with
 leaving them to dry further in a clean place for at
 fr  sufficient product drying. The drying time that is                           fr
 least 20 minutes.  actually required depends directly on parameters
 Once the accessories are dry, inspect and pack   that are the sole responsibility of the user (loading
 them as soon as possible (see the section entitled   configuration, how many items are loaded and how
 "Inspection/maintenance" and "Packaging").  closely together they are loaded, condition of the
 sterilizer, etc.) and must therefore be established   Fig. 9  Début du cycle de mesure  Fig. 10  Section coronaire / médiane
 8.2.3  Inspection/Maintenance  by the user. However, the drying time must never
 be less than 20 minutes.                     Section apicale
 Inspect all accessories once they have been   • The gravity method is less effective and   REMARQUES  Vous pouvez observer la progression de l’instru-
 cleaned or cleaned/disinfected. Dispose of any   must only be used if it is not possible to use the   Si vous n’entendez pas les deux bips et   ment à l’approche de la partie apicale du canal sur
 defective accessories immediately. These defects   fractionated vacuum method.  si l’instrument ne bouge pas, ceci in-  l’agrandissement – Zoom apical (Fig.11).
 include:  • Steam  sterilizer  in  accordance  with   dique que la connexion est mauvaise :
 • Deformation of the plastic  DIN EN 13060 or DIN EN 285, ANSI/AAMI ST 79  •  Débranchez le câble de mesure du pa-
 • Corrosion  • The sterilization process must be validated   tient.
 • Discoloration of the plastic  in accordance with DIN EN ISO 17665 (valid   •  Vérifiez le branchement approprié entre
 installation and operational qualification (IQ and
 Accessories that are still contaminated must be   OQ) and product-specific performance qualification   le câble de mesure et les accessoires.
 cleaned and disinfected again. Maintenance is not   (PQ))  •  Nettoyez la zone de contact de la pince
 required. Do not use instrument lubricant.   • Maximum sterilization temperature 135 °C    instrumentale/sonde fourche.
 (275 °F) plus tolerance in accordance with    •  En cas de besoin, irriguez le canal puis
 8.2.4  Packaging   ISO DIN EN ISO 17665  recommencez.
 • Sterilization time (exposure time at sterilization
 Pack accessories into single-use sterilization   temperature) at least 3 minutes at 134 °C (273 °F)
 packaging (single-layer packaging) that fulfils the   Rapid  sterilization  and  sterilization  with  AVERTISSEMENTS  Fig. 11  Section apicale - bleu
 following requirements:  unpackaged accessories are not permitted.  •  Nous vous recommandons de ne pas
 • Complies with DIN EN ISO/ANSI AAMI ISO 11607  Dry heat sterilization, radiation sterilization and   continuer  la  mesure  si  vous  n’entendez   La ligne qui indique la position exacte de l’instru-
 • Suitable for steam sterilization (withstands   sterilization using formaldehyde, ethylene oxide or   pas les deux signaux sonores.  ment dans la section apicale change de couleur en
                                              fonction de sa position ; bleu, vert puis jaune (Fig.
 temperatures of up to 142 °C (288 °F) or more,   plasma are also not permitted.  •  Reportez-vous au menu Réglages et   12, Fig. 13).
 sufficient vapour permeability)  sélectionnez le mode CHECK pour vérifier
              que l’appareil fonctionne correctement.
 WARNINGS
 •  Regional requirements have to be
 followed additionally.  7.4.3  Localisation de l’apex
          Section coronaire et médiane
          Introduisez doucement l’instrument dans le canal.
          La progression de l’instrument des sections coro-
          naires et médianes jusqu’à la région apicale, est
          représentée sur l’image du canal complet par une
          ellipse qui se déplace en continu vers le bas
          (Fig. 10).
                                              Fig. 12  Section apicale - vert
                                 Endo Easy Efficient ®                                                                 #                                                               26306-EYAL - 26306-EYAL | 2 - B | 18-08-14 | 12:05:15 | SR:-- | Yellow   26306-EYAL - 26306-EYAL | 2 - B | 18-08-14 | 12:05:15 | SR:-- | Cyan   26306