Page 176 - fif
P. 176

25567_eyal-Y-MK1-837 - 25567_pnim_y-mk1-837_eyal.1.1 | 7 - B | 18-06-11 | 09:03:44 | SR:-- | Cyan
  #25567_eyal-Y-MK1-837 - 25567_pnim_y-mk1-837_eyal.1.1 | 7 - B | 18-06-11 | 09:03:44 | SR:-- | Black
  25567_eyal-Y-MK1-837 - 25567_pnim_y-mk1-837_eyal.1.1 | 7 - B | 18-06-11 | 09:03:44 | SR:-- | Yellow
  25567_eyal-Y-MK1-837 - 25567_pnim_y-mk1-837_eyal.1.1 | 7 - B | 18-06-11 | 09:03:44 | SR:-- | Magenta
 #
       Appendix 2

               ELECTROMAGNETIC COMPATIBILITY                                                       Guidance and manufacturer’s declaration – electromagnetic immunity for not life
               EMC according to IEC 60601-1-2 ed.4                                                                           support equipment
                                                                                                  The Product is intended for use in the electromagnetic environment specified below.
                           ®
               The ProPex  II is intended for use in professional healthcare facility or          The customer or the user of the Product should assure that it is used in such an
               home healthcare electromagnetic environment specified in this chapter.             environment.
                                                  ®
               The customer or the user of the ProPex  II should assure that it is used             Immunity   IEC 60601   Compliance   Electromagnetic environment - guidance
               in such an environment.                                                               tests      Test level   level
                                                   ®
               Changes or modifications to this ProPex  II not expressly approved by                                                  Portable and mobile RF communications
                                                                                                                                      equipment should be used no closer to
               the manufacturer may result in increased emissions or decreased                                                        any part of the Product, including cables,
                                                 ®
               immunity performance of the ProPex  II and could cause EMC issues                                                      than the recommended separation
                                                    ®
               with this or other equipment. This ProPex  II is designed and tested to                                                distance calculated from the equation
               comply with applicable regulations regarding EMC and shall be installed                                                applicable to the frequency of the
               and put into service according to the EMC information stated as follows.                                               transmitter.
                                                                                                                                      Recommended separation distance:
               Use of portable phones or other radio frequency (RF) emitting                                                          d = 0.35 √P    80 MHz to 800 MHz
                                         ®
               equipment near the ProPex  II may cause unexpected or adverse                                                          d = 0.7 √P    800 MHz to 2.5 GHz
               operation.                                                                                                             Where P is the maximum output power
                           ®
               The ProPex  II must not be used adjacent to, or stacked with, other                Radiated RF   3 V/m     3 V/m       rating in the transmitter in watts (W)
                                                                        ®
               equipment. If adjacent or stacked use is necessary, the ProPex  II                 IEC         80 MHz to   80 MHz to   according to the transmitter
                                                                                                                                      manufacturer and d is the recommended
               must be tested to verify normal operation in the configuration in which            61000-4-3    2.5 GHz    2.5 GHz     separation distance in metres (m).
               it is being used.
                                                                                                                                      Field strengths from fixed RF
                                                                                                                                      transmitters, as determined by an
                                                                                                                                                            a
                                                                                                                                      electromagnetic site survey  , should be
                                                                                                                                      less than the compliance level in each
                                                                                                                                                    b
                                                                                                                                      frequency range  .
                                                                                                                                      Interference may occur in the vicinity of
                                                                                                                                      equipment marked with the following
                                                                                                                                      symbol:











       176/184                          B EU PIIS DFU PRT / Rev.04 / 02-2018 (Old ZF 1902127.X)  B EU PIIS DFU PRT / Rev.04 / 02-2018 (Old ZF 1902127.X)           173/184