                  Use of non-original cables or accessories may result in increased
 Guidance and Manufacturer's declaration - Electromagnetic Immunity
                                                                         ®
                  emissions or decreased immunity performance of the ProPex  II.
 The Product is intended for use in the electromagnetic environment specified below.
 The customer or the user of the Product should assure that it is used in such an
 environment.
                  The table below lists cables and accessories for which the manufacturer
 Immunity   IEC 60601  Compliance level  Electromagnetic   claims EMC compliance:
 tests  Test level  environment - guidance
 Floors should be wood,
 Electrostatic   ± 6 kV contact  ± 6 kV contact  concrete or ceramic tile.   Description  Details
 discharge   If floors are covered with   Measurement cable  Original only.
 (ESD) IEC   ± 8 kV air  ± 8 kV air  synthetic material, the   Accessories:
 61000-4-2  relative humidity should
 be at least 30%.  Lip Clip    Original only.
 Power   Power frequency   Connection Hook  Original only.
 frequency   magnetic fields should be   Connection Fork  Original only.
 (50/60 Hz)   3 A/m  3 A/m  at levels characteristic of
 magnetic   a typical location in a   Original Switching charger only:
 field IEC   typical commercial or   Charger  Input: 100 - 240 VAC, 50 - 60 Hz
 61000-4-8  hospital environment.  Output: 6 VDC ± 5%, 1000 mA
 NOTE: U  is the a.c. mains voltage prior to application of the test level.