Page 124 - Basic Monitoring in Canine and Feline Emergency Patients
P. 124

atmospheric gases in the mask in addition to the
            fresh gas flow into the mask. This dilutes the con-
  VetBooks.ir  centration of CO  and leads to a falsely low
                           2
            ETCO  as compared to an intubated patient.
                 2
            However, knowing that there is CO  dilution, face
                                        2
            mask–derived ETCO  readings can still be trended
                             2
            over time within a single patient and provide clini-
            cal information. It is noteworthy to mention that
            some capnography systems  (like the Oridion sys-
            tems) have a dual nasal and mouth sampling device
            that  may  give  a  better  idea  of  real  ETCO   in
                                                 2
            patients who are panting/breathing through their
            mouth.
              Other advantages of sidestream capnographs
            include faster warm-up time than mainstream
            units, low dead space added to the circuit, and the
            capability to measure several respiratory gases at
            once including anesthetic gases. The line and adap-
            tor that attach to the ETT are also lightweight
            which reduces the chance of kinking or disconnec-
            tion of the system.
              One other aspect of the sidestream capnograph
            that needs to be mentioned is that condensation
            from humidified gas and patient’s secretions  may
            develop and it can accumulate in the sampling line.
            This can lead to inaccurate ETCO  readings as it
                                        2
            can affect the sampling cell function or sometimes
            cause occlusion of the sampling line. Frequent
            replacement of the sampling line may be required.
            To avoid this or to at least decrease the frequency
            of sampling line replacement, most units come with
            a water trap (see Fig. 6.7) located at the end of the
            sampling tube. Some units also have water perme-
                                 ®
            able tubing such as Nafion  tubing. This tubing has   Fig. 6.7.  Water trap sidestream capnograph. This is a
            a unique property to transfer moisture from one   sidestream capnograph and the water trap can be seen
            side of the membrane to the other.  This is done   on the side of the machine.
            chemically  via  differential  vapor  pressure  (the
            water molecules move from a higher pressure to a   period between the mouth and the detection cham-
            lower pressure) and is therefore very selective for   ber than traditional sidestream technology and thus
            water molecules only. If there is high moisture pres-  a capnograph curve that is almost synchronized
            sure  in  the  tubing,  the  water  molecules  will  be   with the passage of air at the mouth. In addition,
            transferred outside of the tubing and the gas that   the microstream system has less dead space than
            reaches the analysis chamber will be moisture-free   traditional units (less than 0.5 mL of dead space),
            and will not interfere with the CO  reading.  which makes this a useful ETCO  monitor for the
                                       2
                                                                                   2
              A microstream system has also been developed   small-sized  patient.  Other  benefits  of  the  micros-
            which is a variation of the traditional sidestream   tream technology include a low-flow 50 mL/min
            units. In the microstream technology there is no sen-  sample rate, no dilution with supplementary oxy-
            sor at the airway. It uses laser-based molecular   gen, and no cross reaction with other gases. Using a
            spectroscopy as the infrared emission source. The   low-flow sample rate is especially beneficial for our
            expired CO  travels along a thin tube before reach-  small-sized patients which tend to have a low tidal
                     2
            ing the chamber where the detector is located. The   volume  and  a  fast  respiratory  rate.  Using  a  high-
            microstream system allows for a smaller transition   flow  sampling  rate  (i.e. 150  mL/min used  with


             116                                                                         L.A.M. Ilie
   119   120   121   122   123   124   125   126   127   128   129