Page 85 - Basic Monitoring in Canine and Feline Emergency Patients
P. 85

Pulse
  VetBooks.ir                                                      Arterial blood

                                   Absorption                      Venous blood




                                                                   Tissue



                                             Time
             Fig. 4.5.  Light passing through tissues will be absorbed by hemoglobin (Hb) in multiple compartments. By analyzing
             the ratio of absorption between the signal that varies with pulsatile flow versus the signal that is static, the pulse
             oximeter differentiates between the saturation of Hb in arterial blood and other tissues and reports only the saturation
             of the Hb in arterial blood. Therefore, detection of pulsatile flow is critical to the accuracy of the reading.






                          LED



                                                                   Fig. 4.6.  Transmission pulse oximetry
                Sensor                                             probe (Massimo, Irvine, California, USA).
                                                                   LED, light emitting diode.

             Transmission probes are generally poorly tolerated       LED
             for long-term use unless the patient is anesthetized or
             moribund since they fall off easily with movement.
             Also, transmission probes will gradually compress
             the tissue bed and can compress the vessels, reducing
             the all-important arterial blood flow and causing the
             probe to not give a reading. When using a transmis-    Sensor
             sion probe, the LED and sensor portions of the probe
             must be in correct alignment or the signal can be
             distorted, resulting in inaccurate measurements.  Fig. 4.7.  Reflectance pulse oximetry probe (Massimo,
                                                                                       .
               Reflectance probes (Fig. 4.7) have the LED and   Irvine, California, USA). LED, light emitting diode.
             sensor on the same side of the probe. The emitted
             light is reflected off the bone and is then received
             by the sensor. To allow for proper reflection back   probe in order to allow the emitted light to reflect
             to the sensor, the reflectance probe must either be   back to the sensor.
             placed over a flat bone (e.g. the forehead in humans   In anesthetized animals, reflectance probes can
             or the ventral aspect of the first coccygeal bone in   be used on the hard palate or within the esophagus.
             animals) or held at just the right angle to a more   In awake animals, these probes can be used on the
             curved bone to allow proper reflection of the emit-  underside of the base of the tail, over the femoral
             ted light back to the sensor. The distance from the   artery/proximal femoral bone, on the dorsal sur-
             LED to the bone must not be any farther than the   face  of  the  metatarsus,  or  the  plantar side  of  the
             distance between the two sides of the transmission   carpus. Versions of reflectance probes made to be


             Pulse Oximetry                                                                   77
   80   81   82   83   84   85   86   87   88   89   90