Page 89 - Basic Monitoring in Canine and Feline Emergency Patients
P. 89

Table 4.2.  Possible causes of inaccurate pulse oximetry readings.

  VetBooks.ir  Abnormality         Example                   Effect on SpO  reading
                                                                       2
                                   Patient in recent house fire
              Carboxyhemoglobin
                                                             Spuriously high readings
              Methemoglobin        Exposure to toxins or drugs    Spurious readings in the 80s
                                     (see Table 4.1)
              Vasoconstriction     Hypothermia, shock,       Reduced pulsatile signal, increased risk of motion
                                     vasoconstrictive drugs   artifacts, variable effects on readings
              Vasodilation         Hyperthermia, SIRS/sepsis  Venous congestion may lead to venous pulsation
                                                              and artifactually low readings as pulsing venous
                                                              blood is misread as arterial
              Excessive motion     Non-compliant patient     If interpreted as pulsation, may cause erroneous
                                                              readings (both falsely high and low)
              Pigmented tissues    Black pigment on gums/lips  Low signal quality as less light will reach the
                                                              sensor (absorbed by pigment)
              Other light sources/electrical  Fluorescent lights, electrocautery  Variable effects if light is absorbed erroneously by
               energy detected by                             sensor; better compensated for in newer units
               photodiode
              Lack of pulsatile blood flow  Cardiac arrest, thrombosis    Other artifacts (e.g. motion) may be interpreted
                                     of tissue bed            as pulsatile flow and give erroneous readings.
                                                              Pulse oximetry should not be considered
                                                              accurate in low flow/no flow states
              Anemia               Blood loss, hemolysis     Generally accurate except at extremely low
                                                              (<75%) saturation or anemic shock resulting in
                                                              vasoconstriction
              Intravenous dyes     Methylene blue            Spuriously low readings, pigment affects amount
                                                              of light reaching the sensor
              Hyperbilirubinemia/icterus  Hemolysis, liver disease  Generally accurate assuming minimal presence of
                                                              other dyshemoglobinemias (e.g. COHb)
              Shifted OxyHb dissociation   Changes in body temperature,   The correlation between SpO  and PaO  will be
                                                                                         2
                                                                                  2
               curve                 acid–base status, hyper/  shifted (see Fig. 4.4); the practitioner must take
                                     hypoventilation          this into account when interpreting SpO  values
                                                                                          2
                                                              and using them to predict PaO 2
              Arrhythmias          Atrial fibrillation, VPCs  Correlate with ECG, but if signal quality is good
                                                              and SpO  reading is stable, generally do not
                                                                     2
                                                              affect accuracy of readings
             COHb, carboxyhemoglobin; ECG, electrocardiogram; OxyHb, oxyhemoglobin; PaO , partial pressure of oxygen in plasma; SIRS,
                                                                2
             systemic inflammatory response syndrome; SpO , pulse oximetry; VPC, ventricular premature contraction.
                                           2
             hypoxemic. In this setting, despite all measured val-  COHb, the pulse oximeter would read 95%, which
             ues being normal in the blood and mucous mem-  is falsely inflated by 15%. In this setting, a PaO
                                                                                                2
             branes appearing bright pink/red, the patient will   MUST be measured to determine the amount of
             still have a significant tissue oxygen derangement.  oxygen in the plasma available to tissues.
               Dyshemoglobinemias, such as MetHb and       MeHb absorbs infrared and red light equally well,
             COHb, are also particularly confounding to pulse   which is more infrared light than either OxyHb or
             oximetry. The two wavelengths of light emitted in   DeOxyHb absorb. It also absorbs a similar amount
             traditional  pulse  oximetry  cannot  differentiate   of red light to DeOxyHb. Therefore, when there are
             OxyHb from COHb. Pulse oximeters therefore   significant MetHb levels in the blood (typically
             measure COHb as OxyHb and will give spuriously   >30%), this can lead to a pulse oximetry machine
             high readings that do not reflect the true SaO . For   reading in the 80–85% range, regardless of the
                                                2
             example, if a patient has 80% OxyHb and 15%   actual amount of OxyHb.  This might over or

             Pulse Oximetry                                                                   81
   84   85   86   87   88   89   90   91   92   93   94