Page 87 - scales manual v1014 web
P. 87
UWE

UWE

The following outlines the dealer calibration procedure for the majority of UWE models. Some models
use a "TARE" key, others use a "SAMPLE" key. This procedure does not apply to the HGS type - for
this, see below.

1. Switch on the scale while holding down the TARE/SAMPLE key, the display will show F1.
2. Press the MODE key, the offset value or "internal count" will be displayed, this figure should be

adjusted using the trim pot VR1. The correct figure for most models is 10,000 except for the
EBW & GM series models which are set to around 5000.
3. Press the ZERO key, the display will "Zero off".
4. Press the TARE/SAMPLE key, now press the ZERO key, wait a few seconds, press the
TARE/SAMPLE key again, now press the ZERO key.
5. A small arrow or ring should have appeared to the right of the display.
6. Now place the lower calibration weight on the scale, this is usually one third of the capacity of
the scale. (The value of this weight can be determined by following the user calibration
procedure below). The display should be showing a value of around 11,000. (Except EBW &
GM series). Now press the MODE key, the scale will calibrate itself, show "DONE" and return
to the weighing mode.

UWE Quick Calibration

1. Switch the scale on while holding down the MODE key, the display will show "CAL?"
2. Press the MODE key again, the display will show "YES" then "LOAD" and then change by itself

and alternate between two acceptable calibration weights, choose one of these weights and
place it on the scale.
3. The scale will now self-calibrate and return to the weighing mode.

UWE HGS

Switch scale on while holding down the middle key on the top row of buttons, display will show F1

1. Press the left-hand button on the top row, the internal count will now be displayed, set this
figure to 10,000 using VR1 pot.

2. Switch off the scale, now switch on again while holding down the left hand button on the top
row, the display will show "CAL- ZERO", the display will change by itself ( this can take a while
) to show "LOAD-span" followed by a choice of two acceptable calibration weights, choose
your preferred weight and place it on the scale. After a while the display will show done and
return to the weighing mode.