Page 46 - Brachytherapy
P. 46

Source strength






























            Long  half-life  sources  maintained  within  a  permanent




            inventory should be checked at some established frequency for





            their change in source strength (activity) with time. Calculated




            values, often obtained by decaying a prior calibration, should




            be compared with measurements. Short half-life sources, used




            either in temporary or permanent implants, should have their





            activity measured at the time of receipt and compared with the




            manufacturer’s value.