Page 43 - Brachytherapy
P. 43

Constancy check of callibrated
                                                                                                 dosimeter







































                            The constancy of the response of the calibrated dosimeter system may



                            be checked by periodic measurement of a long half-life source, such as



                            Caesium - 137 in the case of a well type chamber. It is necessary to use



                            a  special  source  holder  that  will  position  the  check  source  in  a



                            reproducible  manner,  since  the  ionization  chamber  response  is  very



                            dependent  on  source  position  and  orientation.  This  periodic



                            measurement also provides a good quality assurance check of the entire



                            measuring system.