Page 45 - Brachytherapy
P. 45

Mechanical properties


































                          The  mechanical  integrity  of  a  source  must  be  checked  at  regular



                          intervals by visual inspection, leak testing and activity measurement.


                          Applicators,  because  of  their  repeated  clinical  use,  undergo  severe



                          handling,  cleaning  and  sterilization.  Visual  inspection  and



                          radiographic  evaluation  of  all  applicators  should  be  performed  at


                          some pre-established frequency. For gynaecological applicators it is



                          necessary to check that the assembly is structurally sound, that all



                          clamps, screws and retaining devices are functioning properly and


                          that the source insert carriers seat correctly in the colpostats.
















                                                                                                                                                                                                                                                                            41