Check of source positioning with a







                                       after loading devices















      The position of sources placed within afterloading devices




      can be determined with autoradiographs. It is necessary to



      check that the unit will position the source with millimetre




      accuracy  at  predetermined  programmed  positions  along




      treatment  catheters.  The  use  of  appropriate  radiographic



      markers and combination of a radiographic image with an




      autoradiograph  is  a  convenient  method  for  checking



      source positioning.