Appendix




                   Address Range                  Data
                       49600 to 49663                Phase C Pair 6 2D Data
                       49900 to 49963                Phase A Pair 7 2D Data
                       50200 to 50263                Phase B Pair 8 2D Data
                       50500 to 50563                Phase C Pair 9 2D Data
                       50800 to 50863                Phase A Pair 10 2D Data
                       51100 to 51163                Phase B Pair 11 2D Data
                       51400 to 51463                Phase C Pair 12 2D Data
                   48200 to 51563                 2D Data for the Alert range
                       48200 to 48263                Phase A Pair 1 2D Data
                       48500 to 48563                Phase B Pair 2 2D Data
                       48800 to 48863                Phase C Pair 3 2D Data
                       49100 to 49163                Phase A Pair 4 2D Data
                       49400 to 49463                Phase B Pair 5 2D Data
                       49700 to 49763                Phase C Pair 6 2D Data
                       50000 to 50063                Phase A Pair 7 2D Data
                       50300 to 50363                Phase B Pair 8 2D Data
                       50600 to 50663                Phase C Pair 9 2D Data
                       50900 to 50963                Phase A Pair 10 2D Data
                       51200 to 51263                Phase B Pair 11 2D Data
                       51500 to 51563                Phase C Pair 12 2D Data


              8.2.8.  Hydro PD Card Measurement Details

                   Address     Data                            Notes
                   12000       User Range Phase A Pair 1       The year of the last measurement
                                   Year                            Expressed in 4 digits (i.e., 2019)
                   12001       User Range Phase A Pair 1       The month of the last measurement
                                   Month                           1 = January … 12 = December
                   12002       User Range Phase A Pair 1       The day of the month of the last measurement
                                   Day
                   12003       User Range Phase A Pair 1       The hour of the last measurement
                                   Hour                            Expressed in 24-hour mode (0-23)
                   12004       User Range Phase A Pair 1       The minute of the last measurement
                                   Minute                          0-59
                   12005       User Range Phase A Pair 1       The second of the last measurement
                                   Second                          0-59
                   12006       Measurement Type                This indicates the measurement range where:
                                                                   0 = User Range
                                                                   1 = Auto Range
                                                                   2 = Alert Range
                   12007       User Range Phase A Pair 1       The number of AC cycles in the last
                                   AC Cycles                       measurement
                   12008       User Range Phase A Pair 1       The AC Max Value – for Iris Power use only
                                   AC Max Value





                                                              98