Table 2.  Guidance and manufacturer’s declaration—
              electromagnetic immunity—for all equipment and systems a b
                           IEC 60601 test          Electromagnetic
                Immunity test         Compliance level              English
                              level              environment—guidance
              Electrostatic   ±6 kV contact  ±8 kV contact  The relative
              discharge                          humidity should be
              (ESD) (IEC   ±8 kV air  ±15 kV air  at least 15 percent.
              61000-4-2)
              Electrical fast   ±2 kV for   ±2 kV for   Mains power quality
              transient/  power supply  power supply  should be that of a
              burst (IEC  lines       lines      typical commercial
              61000-4-4)                         or hospital
                          ±1 kV for   ±1 kV for   environment.
                          input/output   input/output
                          lines       lines
              Surge       ±1 kV       ±1 kV      Mains power quality
              (IEC        differential   differential   should be that of a
              61000-4-5)  mode        mode       typical commercial
                                                 or hospital
                          ±2 kV       ±2 kV      environment.
                          common      common
                          mode        mode
                               b
              Voltage     <5% U       <5% U      Mains power quality
                                          T
                               T
              dips, short   (>95% dip   (>95% dip   should be that of a
              interruptions,   in U ) for 0.5   in U ) for 0.5   typical commercial
                                        T
                             T
              and voltage   cycle     cycle      or hospital
              variations on                      environment.
              power supply   40% U     40% U
                               T
                                          T
              input lines   (60% dip   (60% dip
              (IEC        in U ) for 5   in U ) for 5
                                        T
                             T
              61000-4-11)  cycles     cycles
                          70% U       70% U
                                          T
                               T
                          (30% dip in   (30% dip in
                          U ) for 25   U ) for 25
                           T
                                       T
                          cycles      cycles
                          <5% U T     <5% U T
                          (>95% dip in   (>95% dip in
                          U ) for 5 sec  U ) for 5 sec
                           T
                                       T
                                LATITUDE Communicator Patient Manual   89
                                     ®