IEC 60601   Compliance   Electromagnetic environment—
              Immunity test
         English  Radiated   3 V/m  10 V/m   d = 0.5 √P (80 MHz to 1000
                         test level
                                                   guidance
                                    level
                                            MHz)
                        80 MHz to  80 MHz to
              RF (IEC
                                        h
                                 1000 MHz
              61000-4-3)
                                            d = 0.29 √P (380 MHz to 390
                        2.5 GHz
                                            MHz)
                                 3 V/m      d = 0.36 √P (430 MHz to 470
                                 1 GHz to   MHz)
                                 2.7 GHz  i  d = 0.21 √P (800 MHz to 960
                                            MHz)
                                 8 V/m      d = 1.7 √P (1 GHz to 2.7 GHz)
                                 5.1 GHz to   d = 0.21 √P (1.7 GHz to 1.99
                                 5.8 GHz    GHz)
                                            d = 0.21 √P (2.4 GHz to 2.57
                                            GHz)
                                            d = 0.63 √P (5.1 GHz to 5.8 GHz)
                                            where P is the maximum
                                            output power rating of
                                            the transmitter in watts
                                            (W) according to the
                                            transmitter manufacturer
                                            and d is the recommended
                                            separation distance in
                                            meters (m).
                                            Field strengths from
                                            fixed RF transmitters,
                                            as determined by an
                                            electromagnetic site
                                                 f
                                            survey,  should be less
                                            than the compliance level
                                            in each frequency range. g
                                            Interference may occur in
                                            the vicinity of equipment
                                            marked with the following
                                            symbol:
                                                  Non-iodizing
                                                  electromagnetic
                                                  radiation
              a.  The LATITUDE Communicator is intended for use in the electromagnetic
                environment specified above. The customer or the user of the LATITUDE
                Communicator should assure that it is used in such an environment.
              b.  At 80 MHz and 1 GHz, the higher frequency range applies.


              92     LATITUDE Communicator Patient Manual
                          ®