Page 95 - pace maker 359371-001_LATITUDE_Patient_Manual e-magazine
P. 95
IEC 60601 Compliance Electromagnetic environment—
Immunity test
test level level guidance
Conducted 3 Vrms 3 Vrms d = 1.7 √P (150 kHz to 80 MHz) English
RF (IEC 150 kHz
61000-4-6) to 80 MHz 10 Vrms in where P is the maximum
ISM and output power rating of
amateur the transmitter in watts
radio bands (W) according to the
between transmitter manufacturer
150 kHz and and d is the recommended
80 MHz e separation distance in
meters (m).
Field strengths from
fixed RF transmitters,
as determined by an
electromagnetic site
survey, should be less
f
than the compliance level
in each frequency range. g
LATITUDE Communicator Patient Manual 91
®