4.3 Rough-In Secondary Check Unit Assembly (Pressurized Outlets Only)
                   Chapter 4
                   4.3    Rough-In Secondary Check Unit Assembly (Pressurized Outlets
                          Only)
                   Tools required:   Small screwdriver

                   Removal




                                 WARNING:
                          Notify all facility personnel whenever shutting off any medical gas supply.
                          Failure to do so could result in patient injury or death.




                                 WARNING:
                          Shut off the gas supply to the outlet before servicing the rough-in secondary
                          check unit assembly. Failure to do so could result in personal injury or
                          equipment damage.


                          1. Shut off the gas supply to the outlet.

                          2. Remove the finish/frontbody assembly (refer to procedure 4.1).


                          3. Note the removal sequence of the internal parts in the rough-in secondary check
                              unit assembly to ensure proper replacement during the installation procedure.


                          4. Using a small screwdriver, remove the retaining ring (A) from the rough-in
                              assembly (B) (see figure 4-3 on page 4–7).


                          5. Remove the washer (C).

                          6. Using the small screwdriver, carefully remove the rubber bushing (D) from the
                              rough-in assembly (B).

                          7. Remove the secondary check (E).

                          8. Remove the spring (F).





















                   Page 4-6              DiamondCare® Recessed Wall Outlet and Electrical Accessories (MAN01-045)