4.4 Rough-In Assembly (Vacuum Outlets Only)
Chapter 4
4.4 Rough-In Assembly (Vacuum Outlets Only)
Tools required: Small screwdriver
Removal
WARNING:
Notify all facility personnel whenever shutting off any medical gas supply.
Failure to do so could result in patient injury or death.
WARNING:
Shut off the service supply to the outlet before servicing the vacuum outlet
bushing. Failure to do so could result in personal injury or equipment
damage.
1. Shut off the service supply to the outlet.
2. Remove the frontbody assembly (refer to procedure 4.1).
3. Note the removal sequence of the internal parts in the rough-in assembly to
ensure proper replacement during the installation procedure.
4. Using a small screwdriver, remove the retaining ring (A) from the rough-in
assembly (B) (see figure 4-4 on page 4–9).
5. Remove the washer (C).
6. Using the small screwdriver, carefully remove the rubber bushing (D) from the
rough-in assembly (B).
Replacement
1. Replace all the parts included in the replacement kit.
2. To install the replacement rough-in assembly, reverse the removal procedure.
3. To ensure proper operation of the rough-in assembly, perform the “Function
Checks” on page 2–3.
Page 4-8 DiamondCare® Recessed Wall Outlet and Electrical Accessories (MAN01-045)