"Oil-Less” Medical Air Medical Systems
4.0 Troubleshooting
Problem Possible Causes Solution
Failure to start Main power disconnected Turn on main power
Power failure Restore power
Main fuse blown Replace fuse
Fuse blown in control circuit Replace fuse
Overload tripped on starter Reset & check for system overload
High temperature switch activated Allow unit to cool; reset switch & check
for over temperature condition
Pressure switch open Adjust or replace switch
Loose or faulty connection Check & tighten all wire connections
Power failure Main fuse blown Replace fuse
Fuse blown in control circuit Replace fuse
Compressor shuts off Overload tripped on starter Reset & check for system overload
unexpectedly
Pressure switch out of adjustment Adjust or replace
High temperature switch activated Allow unit to cool; reset switch & check
for over temperature condition
Motor overheating Low voltage Check for proper supply voltage
V-belt too tight Adjust belt tension
Defective motor Contact BeaconMedæs
Compressor runs hot Wrong rotation Switch motor leads
Incorrect pressure setting Adjust pressure switch
Faulty compressor valves Contact BeaconMedæs
Check if valve or line to receiver Replace if necessary
is leaking or plugged
Intake filter clogged Clean or replace
4-1