Troubleshooting Guide



                      Symptom                    Possible Cause                  Corrective Action
             13)Pressure or vacuum digital  a. Incorrect digital display “unit  a. Use digital display module test mode
                display does not match        of measure”.                  as described on page 41 to check
                pressure or vacuum levels in  b. Faulty digital display module.  current unit of measure (PSI, in Hg,
                pipeline.                   c. Faulty sensor module.        kPa, mm Hg).  If incorrect,
                                                                            reprogram unit of measure.  Refer to
                                                                            Set-Up instructions pages 35
                                                                            through 40.
                                                                          b. If alarm panel contains another
                                                                            working digital display module,
                                                                            remove field wiring connector from
                                                                            suspect module and connect to
                                                                            working (test) module.  If test
                                                                            module provides correct reading,
                                                                            replace faulty digital display module.
                                                                          c. If test module in step b above also
                                                                            gives incorrect display, replace faulty
                                                                            sensor module.

             14)Low alarm with -- FF II -- display  No pressure or vacuum  a. Correct sensor wiring.  If sensor is
                on digital display module.  signal to digital display       mounted remotely, verify field wiring
                                            module as a result of:          continuity and polarity from sensor
                                            a. Sensor wiring disconnected   to digital display module connector
                                              or wiring polarity reversed.  pins 5 (+) and 6 (-).
                                            b. Faulty digital display module. b. If alarm panel contains another
                                            c. Faulty sensor module.        working digital display module,
                                                                            remove field wiring connector from
                                                                            suspect module and connect to
                                                                            working (test) module.  If test
                                                                            module provides any reading other
                                                                            than -- FF II -- , replace faulty digital
                                                                            display module.
                                                                          c. If test module in step b above also
                                                                            gives -- FF II --, replace faulty sensor
                                                                            module.

             15)Low alarm with -- FF 22 -- display  Pressure or vacuum signal  a. Apply appropriate pressure or
                on digital display module.  to digital display module out   vacuum to pipeline.
                                            of measurement range (too     b. If alarm panel contains another
                                            low) as a result of:            working digital display module,
                                            a. No pressure or vacuum in     remove field wiring connector from
                                              pipeline.                     suspect module and connect to
                                            b. Faulty digital display module.  working (test) module.  If test
                                            c. Faulty sensor module.        module provides any reading other
                                                                            than
                                                                            - - FF 22 --, replace faulty digital display
                                                                            module.
                                                                          c. If test module in step b above also
                                                                            gives -- FF 22 --, replace faulty sensor
                                                                            module.



         48
              Part No. 6-847684-00 Rev. E00