                   Introduction
                   Chapter 1: Introduction

                                 NOTE:
                                 Hill-Rom recommends using a separate pair of wires for each input to the
                                 Master Alarm. Also, even if the network communications option (alarm Web
                                 page accessibility) is not initially selected, run Cat-5 Ethernet cable to each
                                 rough-in assembly. It will cost the facility more to run the cable after
                                 installation.

                                 2. Master Alarm Rough-In—The Master Alarm Rough-In Assembly
                                     provides connections for field wiring to the power supply and incoming
                                     source signals to the Breakout P.C. Board. It is installed prior to the dry
                                     wall installation and finishing.


                                 3. Master Alarm Front Panel—The Master Alarm Front Panel houses the
                                     Master Alarm Module. It consists of the finish trim plate and hinged door
                                     assembly that mounts to the Master Alarm Rough-In Assembly.

                                 4. Master Alarm Module—The Master Alarm Module monitors discrete
                                     inputs from source equipment and; if the Network Communications Board
                                     is installed in the Area Alarm, the Area Alarm Modules. When any errors
                                     or faults occur, it informs the facility’s staff immediately with both audible
                                     and visual indications.

                                 5. Master Alarm Breakout P.C. Board—The Master Alarm Breakout P.C.
                                     Board interfaces with the source equipment field wiring, two dry-contact
                                     relays that can be used as outputs, and one dry-contact relay that can
                                     interface with conventional building automation equipment. This Master
                                     Alarm Breakout P.C. Board is located in the Master Alarm Rough-In.

                                 6. Area Alarm (Local Emergency Alarm)—The Area Alarm contains
                                     between one and six gas-specific Area Alarm Modules that monitor and
                                     display the actual pressure or vacuum of gas lines within the facility. The
                                     Area Alarm consists of the Area Alarm Front Panel and the Area Alarm
                                     Rough-In.
                                     There is one dry-contact relay in the Area Alarm that provides an output
                                     signal. The output can be wired to either the normally-closed or normally-
                                     open contacts. It is located on the Area Alarm Breakout P.C. Board, and
                                     marked MAIN ALARM. For wiring diagrams, refer to the foldouts at the
                                     rear of this manual.
                                   • MAIN ALARM RELAY

                                      •   When the Master Alarm is in alarm condition, this relay activates.

                                      •   When the Master Alarm is silenced, this relay stays activated.

                                      •   When the Master Alarm condition is remedied, this relay deactivates.