                   Features
                   Chapter 1: Introduction

                                 The features of the Area Alarm Modules (DS) are as follows:

                                      •   Display (BH)—Displays pressure and error codes.

                                      •   Low Pressure LED (AZ)—Lights a red LED when the pressure or
                                          vacuum is lower than the setpoint.

                                      •   Normal Pressure LED (BA)—Lights a green LED when the
                                          pressure is normal.

                                      •   High Pressure LED (BB)—Lights a red LED when the pressure is
                                          higher than the setpoint.


                                      •   Labels (BG)—Show which type of gas the Area Alarm Module is
                                          monitoring.

                                      •   Increment key (BC)—If the module is in monitoring mode, press the
                                          Increment key to display the high setting. If the module is in
                                          configuration mode, press it to scroll up the menu.

                                      •   Decrement key (BD)—If the module is in monitoring mode, press
                                          the Decrement key to display the low setting. If the module is in
                                          configuration mode, press it to scroll down the menu.

                                      •   Silence key (BF)—Silences the audible alarm.

                                      •   Test key (BE)—If the module is in monitoring mode, press the Test
                                          key to start the self-diagnostic routine for the Area Alarm Module. If
                                          the module is in configuration mode, press it to view current settings.

                                 The rough-in assembly is installed prior to the dry wall installation. The front
                                 panel mounts to the rough-in assembly after the dry wall is finished.

                                 NOTE:
                                 The MedPlus TotalAlert™ Alarm Network Conversion Kit is shipped with a
                                 frame. The frame is surface-mounted to the dry wall around the existing rough-
                                 in box. Then, the front panel mounts to the frame.

                                 They are both shipped at the same time; however, the front panel is in its own
                                 carton to provide better protection for the electronic components and for safe
                                 storage until it is installed.















                   Page 1 - 18           MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)