                   Specifications
                   Chapter 1: Introduction


                   Specifications


                                 For MedPlus TotalAlert ™ Alarm Network specifications, refer to table 1-10
                                 on page 1-40 through table 1-15 on page 1-41.


                                 Refer to table 1-16 on page 1-42 for the types of existing alarms that can be
                                 converted to a MedPlus TotalAlert™ Alarm Network—Master Alarm,
                                 MedPlus TotalAlert™ Alarm Network—Area Alarm, or MedPlus TotalAlert™
                                 Alarm Network—Combination Alarm. Also included in the table are the sizes
                                 that the existing rough-in boxes need to be, and the conversion kits that replace
                                 those existing alarms.


                                                      Table 1-10.  Power Specifications

                              Facility Power Specifications                           Ratings
                    Supply voltage for single power supply alarms       120 V AC, 60 Hz

                    Breaker for single power supply alarms              15 A
                    Supply voltage for dual power supply alarms         120 V AC, 60 Hz
                    Breaker for dual power supply alarms                15 A
                    Field power supply wiring                           12 gauge

                    Fuse rating for internal power supply (except for the  0.5 Amp Slo-Blo® a
                    Master Alarm plus Six Area Alarm Modules)
                    Fuse rating for internal power supply on the Master  0.75 Amp Slo-Blo®
                    Alarm plus Six Area Alarm Modules

                       a. Slo-Blo® is a registered trademark of Littelfuse, Inc.



                                   Table 1-11.  Cabling Requirements from the Area Alarm Module to the
                                                                 Transducer

                                      Specifications                                Dimensions

                    Distance                                            Maximum 4500' (1372 m)
                    Cable                                               Stranded, 22-gauge, shielded, twisted-
                                                                        pair, three-conductor cable
                    Signal Power                                        15 V DC, 75 mA maximum














                   Page 1 - 40           MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)