PROPRIETARY AND CONFIDENTIAL DRAFT 28/Feb/03


                                                                            Detailed Component Descriptions
                                                                                                                 1
                                                                                     Chapter 1: Introduction

                              NOTE:
                              Adapters are needed for the following gas types: vacuum, helium, nitrogen,
                              waste anesthetic gas disposal (WAGD), carbon dioxide, and carbon
                              dioxide/oxygen mixture.


                              The local-mount transducer can only be used with the MedPlus TotalAlert™
                              Alarm Network Conversion Kits (see table 1-16 on page 1-42).



                              WARNING:
                              Do not remove the threaded insert from the transducer. Doing so may
                              result in attaching the transducer to the incorrect gas or vacuum type,
                              endangering patient welfare.

                              Never remove the threaded insert (DW) from the transducer; it is not a
                              serviceable component. If the transducer (DT) needs serviced or calibrated,
                              send both the transducer and the threaded insert (DX) back to the factory.





















































               MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)     Page 1 - 39