Detailed Component Descriptions
Chapter 1: Introduction
The Area Alarm Modules have gas-specific, color-coded labels for both USA
and ISO (see table 1-9 on page 1-34).
Table 1-9. Area Display Module Gas Labels
Background
Gas name Text Color Use
Color
Oxygen White Green USA
Air Black Yellow USA
Vacuum Black White USA
Nitrogen White Black USA/ISO
Nitrous Oxide White Blue USA/ISO
Evacuation (WAGD) White Purple USA
Carbon Dioxide Black Gray USA/ISO
Helium White Brown USA/ISO
CO /O 2 Gray/green White USA
2
Air a Black and White and ISO
White Black
Oxygen Green White ISO
Vacuum Black Yellow ISO
Air a Black and White and French
White Black
Oxygene Green White French
Protoxyde D’Azote White Blue French
(N O)
2
Succion (vacuum) Black Yellow French
Azote (N ) White Black French
2
a. Text is both black and white in a striped fashion.
An additional label space is provided at the base of each module, for field use,
to note the location that the module is monitoring.