Page 628 - Med Plaza and Cancer Center
P. 628

PROPRIETARY AND CONFIDENTIAL DRAFT 28/Feb/03


                                                                            Detailed Component Descriptions
                                                                                                                 1
                                                                                     Chapter 1: Introduction

                              There is one dry-contact relay featured in the Area Alarm that provides an
                              output signal. The output can be wired to either the normally-closed or
                              normally-open contacts. It is located on the Area Alarm Breakout P.C. Board,
                              and marked MAIN ALARM. For wiring diagrams, refer to the foldouts at the
                              rear of this manual.

                               • MAIN ALARM RELAY

                                   •  When the Master Alarm is in alarm condition, this relay activates.

                                   •  When the Master Alarm is silenced, this relay stays activated.

                                   •  When the Master Alarm condition is remedied, this relay deactivates.

                              When the battery needs changed, Real Time clock battery low displays on the
                              Area Alarm’s Home page. The buzzer does not sound. This message logs into
                              the Error Log when the CheckSum Test is run, which is a test that the Area
                              Alarm runs automatically every 24 hours.


                              The Silence key silences the audible alarm. You can set the time between the
                              audible alarms to resound never, every 30 minutes, every 60 minutes, every 90
                              minutes, or every 120 minutes. The alarm code remains on the display, the red
                              LED remains steadily lit, and the reading continues to flash indicating that the
                              alarm condition has not been remedied.


                              NOTE:
                              To comply with the requirements of the City of New York Department of
                              Buildings; Materials and Equipment Acceptance Division (M.E.A.), do not set
                              the alarm return time to “never.” Set the alarm to return every 30 minutes.

                              A Test key starts a self-diagnostic routine.

                               • The seven segment displays light each individual segment.
                               • The LEDs light.

                               • The audible alarm sounds.























               MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)     Page 1 - 31