1
Chapter 1: Introduction
Local Alarm
The Local Alarm is a stand-alone component of the MedPlus TotalAlert™
Alarm Network System.
The Local Alarm provides connections for monitoring up to 16 signals from
source equipment (see figure 1-15 on page 1-35).
Figure 1-15. Local Alarm
There are individual LED alarm indicators for each signal. When an alarm
occurs, the green Normal LED changes to light red, and the alarm sounds.
A Test key tests the LEDs and the audible alarm.
The Alarm Silence key silences the audible alarm. Following an alarm silence,
the red LED remains steadily lit to indicate that the alarm condition has not
been remedied. Each new alarm condition resounds the alarm.
MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra) Page 1 - 35