Page 632 - Med Plaza and Cancer Center
P. 632

PROPRIETARY AND CONFIDENTIAL DRAFT 28/Feb/03


                                                                            Detailed Component Descriptions
                                                                                                                 1
                                                                                     Chapter 1: Introduction

                              Local Alarm

                              The Local Alarm is a stand-alone component of the MedPlus TotalAlert™
                              Alarm Network System.


                              The Local Alarm provides connections for monitoring up to 16 signals from
                              source equipment (see figure 1-15 on page 1-35).


                                               Figure 1-15.  Local Alarm






































                              There are individual LED alarm indicators for each signal. When an alarm
                              occurs, the green Normal LED changes to light red, and the alarm sounds.

                              A Test key tests the LEDs and the audible alarm.


                              The Alarm Silence key silences the audible alarm. Following an alarm silence,
                              the red LED remains steadily lit to indicate that the alarm condition has not
                              been remedied. Each new alarm condition resounds the alarm.














               MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)     Page 1 - 35