                   Detailed Component Descriptions
                   Chapter 1: Introduction

                                 Network Communications Board

                                 The Master Alarm and Area Alarm are stand-alone units unless the optional
                                 Network Communications Board is installed in the Area Alarm. In that case;
                                 you can browse, download, configure, and troubleshoot Master Alarms and
                                 Area Alarms with a personal computer that is connected to the facility’s
                                 Ethernet. The alarm’s Web pages are built-in; no setup is required.

                                 NOTE:
                                 The Area Alarm Module High and Low setpoints which can be configured
                                 only by using their front panels.


                                 When an alarm condition occurs, each Master Alarm can send electronic
                                 messages to up to five users. In addition, the entire MedPlus TotalAlert™
                                 Alarm Network can be remotely monitored over the Internet.

                                 System Power Supply


                                 Each rough-in assembly or back box assembly for the original version of the
                                 MedPlus TotalAlert™ Alarm Network has its own power supply, which is
                                 installed at the factory.

                                 The power supply for the MedPlus TotalAlert™ Alarm Network Conversion
                                 Kits is a single sub-assembly that is installed in an existing rough-in box.


                                 Each power supply converts the facility’s 120 V AC 60 Hz power to 20 V AC
                                 to operate the alarm. This unit contains the main power switch, transformer,
                                 main fuse, and terminal block.

                                 NOTE:
                                 The Local Alarm is not available as a conversion kit.


                                 Transducers

                                 Each transducer is factory-programmed for the specific type of gas or vacuum
                                 shown on its housing. Transducers convert the gas or vacuum pressure into a
                                 digital signal, then they send that signal to the corresponding Area Alarm
                                 Module. If the transducers are not connected to the appropriate gas-specific
                                 Area Alarm Module, an error message (A07) displays on the Area Alarm
                                 Module. The transducers are connected to Area Alarm Modules with stranded,
                                 twisted-pair, shielded cable (two shielded wires). To ensure dependable
                                 operation, the cable should not be greater than 4500' (1372 m) in length.










                   Page 1 - 36           MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)