PROPRIETARY AND CONFIDENTIAL DRAFT 28/Feb/03

                                                                                           Function Checks
                                                                      Chapter 2: Troubleshooting Procedures


               Function Checks


                              If you are troubleshooting a problem, perform“Initial Actions” on page 2-4
                              before running the self-tests.
                                                                                                                 2
                       Master Alarm Self-Test


                              The Master Alarm, Area Alarm, and Local Alarm have self-testing routines.

                              The Master Alarm routine tests the following:
                                   •  Vacuum Fluorescent Display (VFD)

                                   •  LEDs

                                   •  Audible alarm

                              It also displays the following:
                                   •  Flash memory checksum

                                   •  IP address
                                   •  Programmed alarm setup

                                   •  LED mapping

                              During this routine, the buzzer and the main relay are disabled; however, the
                              auxiliary relay still triggers when an alarm condition occurs.

                              1. To start the test, press and release the Test key.

                              2. 0 through 9 and A through Z show in rotation for all of the places on the
                                 display.

                              3. LED Test displays.

                              4. The LEDs go out.


                              5. LEDs 1 through 9 individually light green.

                              6. LEDs 1 through 9 individually light red.

                              7. Audible Test displays.


                              8. The alarm sounds for 2 seconds.









               MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)      Page 2 - 5