                   Error Codes and Troubleshooting Tips
                   Chapter 2: Troubleshooting Procedures


                   Error Codes and Troubleshooting Tips


                          Area Alarm


                                 The error codes for the Area Alarm Modules are used to help identify and
                                 solve problems with the facility’s gas system or the MedPlus TotalAlert™
                                 Alarm Network (see table 2-1 on page 2-8).


                                                 Table 2-1.  Area Alarm Module Error Codes

                    Error Code             Error Description                  Possible Corrective Action
                    A 01         Transducer pressure below sensor range. Turn on the shutoff valves, or check
                                                                          the source equipment. Go to RAP
                                                                          2.1 on page 2-13.
                    A 02         Transducer pressure above sensor range. Check the source equipment, or
                                                                          replace the transducer. Go to RAP
                                                                          2.2 on page 2-13.
                    A 03         Transducer communication time out.       Connect the transducer, or replace
                                                                          the transducer. Go to RAP 2.3 on
                                                                          page 2-14.

                    A 04         Area Alarm Module RAM error.             Replace the Area Alarm Module.
                                                                          Go to RAP 2.9 on page 2-18.

                    A 05         Area Alarm Module ROM error.             Replace the Area Alarm Module.
                                                                          Go to RAP 2.10 on page 2-19.

                    A 06         Transducer status fault.                 Replace the transducer. Go to RAP
                                                                          2.4 on page 2-15.

                    A 07         Incorrect transducer connected to Area   Install the correct gas-specific
                                 Alarm Module.                            transducer. Go to RAP 2.5 on page
                                                                          2-16.
                    A 08         Area Alarm Module programmed as          Install the correct gas-specific Area
                                 vacuum, but units are not inHg or        Alarm Module. Go to RAP 2.8 on
                                 mmHg.                                    page 2-18.
                    A 09         Area Alarm Module programmed as          Install the correct gas-specific Area
                                 pressure, but are units not in PSI or kPa. Alarm Module. Go to RAP 2.8 on
                                                                          page 2-18.

                    A 10         Transducer is programmed as an invalid  Install the correct gas-specific
                                 type of gas.                             transducer. Go to RAP 2.6 on page
                                                                          2-16.
                    A 11         Transducer short circuit detected.       Replace the transducer wiring. Go
                                                                          to RAP 2.7 on page 2-17.





                   Page 2 - 8            MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)