Page 750 - Med Plaza and Cancer Center
P. 750

PROPRIETARY AND CONFIDENTIAL DRAFT 28/Feb/03

                                                                        Error Codes and Troubleshooting Tips
                                                                      Chapter 2: Troubleshooting Procedures

                       Transducer


                              A green LED inside the transducer’s transparent housing indicates whether the
                              transducer is functioning properly (see table 2-4 on page 2-11).


                                         Table 2-4.  Transducer LED Troubleshooting Codes
                                                                                                                 2
                Error Codes            Possible Cause                      Corrective Action
                 Slow steady  Transducer is functioning          No action is required.
                 flash         properly.
                 Rapid flash   Device error.                     Replace the transducer.
                 LED off       No power to the transducer, or    Connect the power, or replace the
                               the transducer has failed.        transducer.


























































               MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)     Page 2 - 11