Page 750 - Med Plaza and Cancer Center
P. 750
PROPRIETARY AND CONFIDENTIAL DRAFT 28/Feb/03
Error Codes and Troubleshooting Tips
Chapter 2: Troubleshooting Procedures
Transducer
A green LED inside the transducer’s transparent housing indicates whether the
transducer is functioning properly (see table 2-4 on page 2-11).
Table 2-4. Transducer LED Troubleshooting Codes
2
Error Codes Possible Cause Corrective Action
Slow steady Transducer is functioning No action is required.
flash properly.
Rapid flash Device error. Replace the transducer.
LED off No power to the transducer, or Connect the power, or replace the
the transducer has failed. transducer.
MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra) Page 2 - 11