                   2.3  Transducer Communication Error
                   Chapter 2: Troubleshooting Procedures

                   2.3    Transducer Communication Error


                                 1. An Area Alarm Module displays error code A 03.
                                     Yes   No
                                      ↓     → Go to “Final Actions” on page 2-7.


                                 2. Transducer is connected properly.
                                     Yes   No
                                      ↓     → Connect the transducer properly (refer to procedure 4.23 on page
                                                4-62). If you need more information about how to connect the
                                                transducer, refer to the MedPlus TotalAlert™ Alarm Network—
                                                Conversion Kits Installation Instructions (is554). If this solves
                                                the problem, go to “Final Actions” on page 2-7. Otherwise, go to
                                                step 3.


                                 3. Transducer has a steady flash on its heartbeat LED.
                                     Yes   No
                                      ↓     → Replace the transducer (refer to procedure 4.23 on page 4-62). If
                                                this solves the problem, go to “Final Actions” on page 2-7.
                                                Otherwise, go to step 4.

                                 4. If the problem still exists, call Hill-Rom Technical Support at
                                     (800) 445-3720.









































                   Page 2 - 14           MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)