                                                                         2.10  Area Alarm Module ROM Error
                                                                      Chapter 2: Troubleshooting Procedures

               2.10 Area Alarm Module ROM Error


                              1. An Area Alarm Module displays error code A 05.
                                 Yes    No
                                  ↓      → Go to “Final Actions” on page 2-7.
                                                                                                                 2
                              2. Area Alarm Module is connected properly.
                                 Yes    No
                                  ↓      → Connect the Area Alarm Module properly (refer to procedure 4.9
                                            on page 4-26). If this solves the problem, go to “Final Actions”
                                            on page 2-7. Otherwise, go to step 3.

                              3. Area Alarm Module is working correctly.
                                 Yes    No
                                  ↓      → Replace the Area Alarm Module (refer to procedure 4.9 on page
                                            4-26). If this solves the problem, go to “Final Actions” on page
                                            2-7. Otherwise, go to step 4.

                              4. If the problem still exists, call Hill-Rom Technical Support at
                                 (800) 445-3720.


               2.11 Battery Low on Master Alarm


                              1. Battery Low displays on the Master Alarm.
                                 Yes    No
                                  ↓      → Go to “Final Actions” on page 2-7.

                              2. Check the Master Alarm’s battery voltage. It is approximately 2.9 V DC.

                                 Yes    No
                                  ↓      → Replace the Master Alarm’s battery (refer to procedure 4.3 on
                                            page 4-12). If this solves the problem, go to “Final Actions” on
                                            page 2-7. Otherwise, go to step 3.

                              3. If the problem still exists, call Hill-Rom Technical Support at
                                 (800) 445-3720.




















               MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)     Page 2 - 19