Page 761 - Med Plaza and Cancer Center
P. 761

PROPRIETARY AND CONFIDENTIAL DRAFT 28/Feb/03

                   2.14  Service Required
                   Chapter 2: Troubleshooting Procedures

                   2.14 Service Required


                                 1. Service Required displays on the Master Alarm.
                                     Yes   No
                                      ↓     → Go to “Final Actions” on page 2-7.


                                 2. Master Alarm is connected to the source equipment properly.
                                     Yes   No
                                      ↓     → Connect the source equipment to the Master Alarm. Refer to
                                                either the MedPlus TotalAlert™ Alarm Network Installation
                                                Instructions (is550) or the MedPlus TotalAlert™ Alarm
                                                Network—Conversion Kits Installation Instructions (is554), if
                                                needed. If this solves the problem, go to “Final Actions” on page
                                                2-7. Otherwise, go to step 3.


                                 3. If the problem still exists, call Hill-Rom Technical Support at
                                     (800) 445-3720.


                   2.15 Gas or Vacuum Wrong Gas

                                 1. Wrong Gas displays on the Master Alarm.
                                     Yes   No
                                      ↓     → Go to “Final Actions” on page 2-7.

                                 2. The Area Alarm Module, at the location showing on the Master Alarm’s
                                     display, is connected to the correct gas-specific transducer.
                                     Yes   No
                                      ↓     → Replace the transducer with one that is for the correct type of gas
                                                or vacuum (refer to procedure 4.20 on page 4-50). If this solves
                                                the problem, go to “Final Actions” on page 2-7. Otherwise, go to
                                                step 3.

                                 3. Replace the Area Alarm Module with one that is for the correct type of gas
                                     or vacuum of the line that it is monitoring (refer to procedure 4.9 on page
                                     4-26). This solves the problem.
                                     Yes   No
                                      ↓     → Call Hill-Rom Technical Support at (800) 445-3720.


                                 4. Go to “Final Actions” on page 2-7.













                   Page 2 - 22           MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)