2.8 Area Alarm Program Errors
Chapter 2: Troubleshooting Procedures
2.8 Area Alarm Program Errors
1. An Area Alarm Module displays error code A 08 or A 09.
Yes No
↓ → Go to “Final Actions” on page 2-7.
2. Replace the Area Alarm Module (refer to procedure 4.9 on page 4-26).
This solves the problem.
Yes No
↓ → Call Hill-Rom Technical Support at (800) 445-3720.
3. Go to “Final Actions” on page 2-7.
2.9 Area Alarm Module RAM Error
1. An Area Alarm Module displays error code A 04.
Yes No
↓ → Go to “Final Actions” on page 2-7.
2. Area Alarm Module is connected properly.
Yes No
↓ → Connect the Area Alarm Module properly (refer to procedure 4.9
on page 4-26). If this solves the problem, go to “Final Actions”
on page 2-7. Otherwise, go to step 3.
3. Area Alarm Module is working correctly.
Yes No
↓ → Replace the Area Alarm Module (refer to procedure 4.9 on page
4-26). If this solves the problem, go to “Final Actions” on page
2-7. Otherwise, go to step 4.
4. If the problem still exists, call Hill-Rom Technical Support at
(800) 445-3720.
Page 2 - 18 MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)