                   2.8  Area Alarm Program Errors
                   Chapter 2: Troubleshooting Procedures

                   2.8    Area Alarm Program Errors


                                 1. An Area Alarm Module displays error code A 08 or A 09.
                                     Yes   No
                                      ↓     → Go to “Final Actions” on page 2-7.


                                 2. Replace the Area Alarm Module (refer to procedure 4.9 on page 4-26).
                                     This solves the problem.
                                     Yes   No
                                      ↓     → Call Hill-Rom Technical Support at (800) 445-3720.

                                 3. Go to “Final Actions” on page 2-7.


                   2.9    Area Alarm Module RAM Error


                                 1. An Area Alarm Module displays error code A 04.
                                     Yes   No
                                      ↓     → Go to “Final Actions” on page 2-7.

                                 2. Area Alarm Module is connected properly.

                                     Yes   No
                                      ↓     → Connect the Area Alarm Module properly (refer to procedure 4.9
                                                on page 4-26). If this solves the problem, go to “Final Actions”
                                                on page 2-7. Otherwise, go to step 3.

                                 3. Area Alarm Module is working correctly.
                                     Yes   No
                                      ↓     → Replace the Area Alarm Module (refer to procedure 4.9 on page
                                                4-26). If this solves the problem, go to “Final Actions” on page
                                                2-7. Otherwise, go to step 4.


                                 4. If the problem still exists, call Hill-Rom Technical Support at
                                     (800) 445-3720.
























                   Page 2 - 18           MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)