                   2.5  Incorrect Transducer
                   Chapter 2: Troubleshooting Procedures

                   2.5    Incorrect Transducer


                                 1. An Area Alarm Module displays error code A 07.
                                     Yes   No
                                      ↓     → Go to “Final Actions” on page 2-7.


                                 2. Transducer is programmed for the correct type of gas.
                                     Yes   No
                                      ↓     → Connect the proper gas-specific transducer (refer to procedure
                                                4.23 on page 4-62). If you need more information about how to
                                                connect the transducer, refer to the MedPlus TotalAlert™ Alarm
                                                Network—Conversion Kits Installation Instructions (is554). If
                                                this solves the problem, go to “Final Actions” on page 2-7.
                                                Otherwise, go to step 3.


                                 3. Transducer has a slow steady flash on its heartbeat LED.
                                     Yes   No
                                      ↓     → Replace the transducer (refer to procedure 4.23 on page 4-62). If
                                                this solves the problem, go to “Final Actions” on page 2-7.
                                                Otherwise, go to step 4.

                                 4. If the problem still exists, call Hill-Rom Technical Support at
                                     (800) 445-3720.

                   2.6    Transducer Program Errors


                                 1. An Area Alarm Module displays error code A 10.
                                     Yes   No
                                      ↓     → Go to “Final Actions” on page 2-7.


                                 2. Transducer has a slow steady flash on its heartbeat LED.
                                     Yes   No
                                      ↓     → Replace the transducer (refer to procedure 4.23 on page 4-62). If
                                                this solves the problem, go to “Final Actions” on page 2-7.
                                                Otherwise, go to step 3.

                                 3. If the problem still exists, call Hill-Rom Technical Support at
                                     (800) 445-3720.