                   2.12  Ethernet Connection is Lost
                   Chapter 2: Troubleshooting Procedures

                   2.12 Ethernet Connection is Lost


                                 1. Ethernet Lost displays on the Master Alarm.
                                     Yes   No
                                      ↓     → Go to “Final Actions” on page 2-7.


                                 2. The Ethernet is connected to the Master Alarm Module.
                                     Yes   No
                                      ↓     → Connect the Ethernet to the Master Alarm Module (refer to
                                                procedure 4.25 on page 4-68). If this solves the problem, go to
                                                “Final Actions” on page 2-7. Otherwise, go to step 3.

                                 3. The Ethernet is connected to the corresponding hub.
                                     Yes   No
                                      ↓     → Connect the Ethernet to the corresponding hub. Refer to the
                                                wiring diagram of the facility’s Ethernet network. If this solves
                                                the problem, go to “Final Actions” on page 2-7. Otherwise, go to
                                                step 4.

                                 4. The Master Alarm Module is operating properly.

                                     Yes   No
                                      ↓     → Replace the Master Alarm Module (refer to procedure 4.2 on
                                                page 4-10). If this solves the problem, go to “Final Actions” on
                                                page 2-7. Otherwise, call Hill-Rom Technical Support at
                                                (800) 445-3720.

                                 5. Facility’s Ethernet is operating properly.
                                     Yes   No
                                      ↓     → Call the facility’s Information Systems department. If this solves
                                                the problem, go to “Final Actions” on page 2-7. Otherwise, go to
                                                step 6.

                                 6. If the problem still exists, call Hill-Rom Technical Support at
                                     (800) 445-3720.






















                   Page 2 - 20           MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)