2.16 Gas or Vacuum Wiring
Chapter 2: Troubleshooting Procedures
2.16 Gas or Vacuum Wiring
1. A type of gas or vacuum and the word Wiring displays on the Master
Alarm.
Yes No
↓ → Go to “Final Actions” on page 2-7.
2
2. The Area Alarm Module, at the location showing on the Master Alarm’s
display, is connected to its corresponding transducer properly.
Yes No
↓ → Properly connect the Area Alarm Module to its corresponding
transducer (refer to procedure 4.9 on page 4-26 and procedure
4.23 on page 4-62). If this solves the problem, go to “Final
Actions” on page 2-7. Otherwise, go to step 3.
3. The wiring between the Area Alarm Module and the transducer is pinched.
Yes No
↓ → Call Hill-Rom Technical Support at (800) 445-3720.
4. Replace the wiring. If this solves the problem go to “Final Actions” on
page 2-7. Otherwise, call Hill-Rom Technical Support at (800) 445-3720.
2.17 Gas or Vacuum Shorted
1. A type of gas or vacuum and the word Shorted displays on the Master
Alarm.
Yes No
↓ → Go to “Final Actions” on page 2-7.
2. The Area Alarm Module, at the location showing on the Master Alarm’s
display, is connected to its corresponding transducer properly.
Yes No
↓ → Properly connect the Area Alarm Module to its corresponding
transducer (refer to procedure 4.9 on page 4-26 and procedure
4.23 on page 4-62). If this solves the problem, go to “Final
Actions” on page 2-7. Otherwise, go to step 3.
3. Replace the wiring between the Area Alarm Module and the transducer. If
this solves the problem go to “Final Actions” on page 2-7. Otherwise, call
Hill-Rom Technical Support at (800) 445-3720.
MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra) Page 2 - 23