                                                              2.21  Connection Lost With Device: Device Name
                                                                      Chapter 2: Troubleshooting Procedures

               2.21 Connection Lost With Device: Device Name


                              1. Connection lost with device: followed by the device name, displays on the
                                 Master Alarm.

                                 Yes    No
                                  ↓      → Go to “Final Actions” on page 2-7.
                                                                                                                 2

                              2. The Ethernet connections are connected properly.
                                 Yes    No
                                  ↓      → Ensure all of the Ethernet connections are connected properly.
                                            Refer to the facility’s wiring diagram of the Ethernet network. If
                                            this solves the problem, go to “Final Actions” on page 2-7.
                                            Otherwise, go to step 3.


                              3. The Area Alarm is powered on.
                                 Yes    No
                                  ↓      → Turn the power to the Area Alarm on (see “Area Alarm Power-
                                            Up” on page 1-112). If this solves the problem, go to “Final
                                            Actions” on page 2-7. Otherwise, go to step 4.

                              4. Check the LEDs on the Network Communications Board (see “Area Alarm
                                 Power-Up” on page 1-112). The LEDs that indicate whether the alarm is
                                 communicating are operating correctly.
                                 Yes    No
                                  ↓      → Replace the Network Communications Board. If this solves the
                                            problem, go to “Final Actions” on page 2-7. Otherwise, call
                                            Hill-Rom Technical Support at (800) 445-3720.

                              5. Call Hill-Rom Technical Support at (800) 445-3720.































               MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)     Page 2 - 27