                                                           2.24  No LEDs or Displays are Lit on the Area Alarm
                                                                      Chapter 2: Troubleshooting Procedures

                              6. The wiring on the transformer is correct.

                                 Yes    No
                                  ↓      → Wire the transformer correctly. For the wiring diagram, refer to
                                            FO 3-4 at the rear of this document. If this solves the problem,
                                            go to “Final Actions” on page 2-7. Otherwise, go to step 7.
                                                                                                                 2
                              7. The voltage across the two orange lead connections on the breakout board
                                 is approximately 20 V AC.
                                 Yes    No
                                  ↓      → Replace the transformer (refer to procedure 4.5 on page 4-16). If
                                            this solves the problem, go to “Final Actions” on page 2-7.
                                            Otherwise, call Hill-Rom Technical Support at (800) 445-3720.

                              8. Go to “Final Actions” on page 2-7.

























































               MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)     Page 2 - 31