2.25 No LEDs are Lit, and the Display is Off on the Master Alarm
Chapter 2: Troubleshooting Procedures
6. The wiring on the transformer is correct.
Yes No
↓ → Wire the transformer correctly. For the wiring diagram, refer to
FO 3-2 at the rear of this document. If this solves the problem,
go to “Final Actions” on page 2-7. Otherwise, go to step 7.
2
7. The voltage across the two orange lead connections on the breakout board
is approximately 20 V AC.
Yes No
↓ → Replace the transformer (refer to procedure 4.5 on page 4-16). If
this solves the problem, go to “Final Actions” on page 2-7.
Otherwise, call Hill-Rom Technical Support at (800) 445-3720.
8. Go to “Final Actions” on page 2-7.