2.29 No Audible Signal on the Master Alarm
Chapter 2: Troubleshooting Procedures
2.29 No Audible Signal on the Master Alarm
1. When an alarm occurs, the audible indicator works correctly.
Yes No
↓ → Replace the Master Module (refer to procedure 4.2 on page 4-
10). If this solves the problem, go to “Final Actions” on page 2-
7. Otherwise, call Hill-Rom Technical Support at (800) 445- 2
3720.
2. Go to “Final Actions” on page 2-7.
2.30 No Audible Signal on the Area Alarm
1. When an alarm occurs, the audible indicator works correctly.
Yes No
↓ → Replace the Area Alarm P.C. board located on the back of the
front panel (refer to procedure 4.8 on page 4-24). If this solves
the problem, go to “Final Actions” on page 2-7. Otherwise, call
Hill-Rom Technical Support at (800) 445-3720.
2. Go to “Final Actions” on page 2-7.
2.31 No Audible Signal on the Local Alarm
1. The Local Alarm audible indicator does not work.
Yes No
↓ → Go to “Final Actions” on page 2-7.
2. The wiring from the Local Alarm P.C. board on the back of the front panel
to the buzzer is connected.
Yes No
↓ → Connect the buzzer’s wiring (refer to procedure 4.14 on page 4-
38). If this solves the problem, go to “Final Actions” on page 2-
7. Otherwise, go to step 3.
3. The buzzer works properly.
Yes No
↓ → Replace the buzzer. If this solves the problem, go to “Final
Actions” on page 2-7. Otherwise, call Hill-Rom Technical
Support at (800) 445-3720.
4. Go to “Final Actions” on page 2-7.
MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra) Page 2 - 37